Oxygen/Pyrolysis Summary
Oxygen
Pyrolysis
Price
$25
Click here for the Oxygen Quote Calculator
Oxygen
Pyrolysis
Turn Around Time
3-5 Days
Oxygen
Pyrolysis
Sample Requirements
1.5-3 mg
Oxygen
Pyrolysis
Supported Elements
Oxygen
Note: Oxygen determination can only be performed on purely organic samples. Inorganics attack the platinized carbon catalyst and significantly reduce its effectiveness and may yield erroneous results. In addition, phosphorous and some inorganics form oxides that can not be pyrolized under the analytical conditions.
Oxygen
Pyrolysis
Accuracy
+ 2%-5%
Oxygen
Pyrolysis
Compatible Sample Types
Solids, Liquids, Oils
Oxygen
Pyrolysis
Incompatible Sample Types
Gases
Oxygen
Pyrolysis
Interferences
Fluorine or inorganics
Note: Oxygen determination can only be performed on purely organic samples. Inorganics attack the platinized carbon catalyst and significantly reduce its effectiveness and may yield erroneous results. In addition, phosphorous and some inorganics form oxides that can not be pyrolized under the analytical conditions.
Oxygen
Pyrolysis
Sample Preparations
Approximately 2 mg of sample is weighed accurately into a tin sample boat.
Oxygen
Pyrolysis
Method Summary
The Perkin-Elmer 2400 Elemental Analyzer is fitted with an oxygen accessory kit to determine oxygen in organic materials. The analyzer uses pyrolysis to convert the oxygen to carbon monoxide. The carbon monoxide is then separated from the other pyrolozates under steady state conditions, and measured as a function of thermal conductivity. The instrument is calibrated prior to sample analysis with an NIST traceable organic standard. System suitability is confirmed by analyzing an NIST traceable organic standard. The standard must check to within + 0.3% of its theoretical value. All sample analyses are bracketed with a standard that must be within + 0.3% of its theoretical value.